UV-VIS-NIR SPECTROPHOTOMETER/UV-3600 [ -1 ]

Detail information

Equipment information Maker:
㈱島津製作所/SHIMADZU

Specs:
The UV-3600 can handle measurement with highly precise transmittance and reflectance, and uses three detectors to handle a range going from the ultraviolet region to the near-infrared region. Using a high-performance double monochromator makes it possible to attain an ultra-low stray-light level with a high resolution.

[Wavelength range]
185 to 3,300 nm
[Resolution]
0.1 nm
[Wavelength accuracy]
Ultraviolet and visible regions: ± 0.2 nm Near-infrared region: ± 0.8 nm
[Noise]
0.00005 Abs max. (500 nm), 0.00008 Abs max. (900 nm), 0.00003 Abs max. (1,500 nm), determined using RMS values for 1-second responses, 2 nm slit width
[Stray light]
0.00008% max. (220 nm, Nal)
0.00005% max. (340 nm, NaNO2)
[Monochromator]
2 × 2 grating-type double monochromator
[Accessories]
ISR-3100 Integrating Sphere Attachment, 60 mm dia.
This attachment is used for measurement of diffuse/specular reflectance and measurement of transmission of liquid or solid sample


Location:
301 at I2CNER buildingⅡ

Other information: